What Voltage Does A Fume Hood Require?

Quick answer: Most lab fume hoods run on a standard 115-120 volt, single-phase outlet, the same kind you have at home. Some hoods and blowers are built for 208-240 volts instead. The exact voltage and amps for any hood are printed on its nameplate (the rating label on the hood) and on its spec sheet, so check there before you plan the wiring.

  • Standard benchtop hoods: usually 115V. Labconco’s Protector hoods, for example, come as 100-115V/10A or 208-230V/5A models.
  • Explosion-proof hoods: often sold for either 115V or 230V. Hemco lists one of its explosion-proof hoods at 115/230V.
  • The exhaust blower: on a ducted hood, it’s usually a separate unit with its own electrical needs. More on that below.
  • 220V vs. 240V: in the US, they’re the same thing.

Nearly all fume hoods, ducted and ductless, are made to run on one of these two common voltages. The trick is matching the hood to the power you already have.

Electrical Currents for Fume Hoods

Think of voltage like water pressure: the higher the voltage, the harder it pushes electricity (the current) through the wiring to your outlet. Not every wire and outlet is built for the same voltage. Some can handle 240 volts, and others are only meant for 120. Ignore that, and equipment can overheat, get damaged or even start a fire.

You can usually tell a 120-volt outlet from a 240-volt outlet just by looking at it. A 120-volt outlet is the one you plug everything into, from your coffee pot to your phone charger. It has three slots (two flat, one round) and is common throughout North America. A 240-volt outlet is bigger, with three or four slots in different shapes, so a regular 120-volt plug won’t fit it.

120-volt and 240-volt outlet comparison

When are 240-volt Outlets Needed?

Some equipment needs more than 120 volts to do its job. Electric dryers, ovens and ranges are designed for 240-volt outlets. An oven wired to 120 volts only puts out about a quarter of the heat it would on 240 volts. A fume hood built for 240 volts is the same way: on the wrong circuit, it won’t run the way it should. Know what voltage your facility has so your fume hood works properly and does its job. Planning for electricity is a key part of laboratory design and layout. Know the location of your fume hood within your laboratory so the wiring is right where you need it.

What Needs Power on a Fume Hood? The Hood, the Outlets and the Blower

A fume hood setup usually has three separate electrical parts, and they don’t always run on the same voltage.

  • The hood itself: the lights, airflow monitor and alarm. On most hoods these run on 115V, with 230V versions made for labs that need them.
  • The outlets on the hood: many hoods have built-in outlets on the side posts for equipment you use inside the hood. These are typically rated at 20 amps, and many are GFCI-protected (University of Kansas fume hood spec).
  • The exhaust blower: on a ducted hood, the blower is usually a separate unit in the ductwork or on the roof, so it needs its own power. Check the blower’s label too, not just the hood’s. Our guide to choosing the right fume hood blower covers sizing, and how does a fume hood work? explains how the hood and blower move air together.

Many hoods route all their wiring to one wiring box, where your electrician makes the final hookup. That’s why it pays to hand your electrician the spec sheet before the hood shows up. Still picking a hood? Start with what size fume hood do I need?, since bigger hoods can come with more lights and outlets to power.

Accommodating Your Fume Hood for a 240V Outlet

If your fume hood is built for 240 volts, make sure the outlet or circuit matches. A 240-volt hood on a 120-volt circuit won’t perform the way it should. A 120-volt hood should never be connected to a 240-volt supply, because it can burn out the motor and electronics. The good news: the plugs are shaped differently, so they won’t fit the wrong outlet. If your hood and your outlet don’t match, don’t force it or grab an adapter. Have a licensed electrician add the right circuit, or ask us about a hood that matches the power you already have.

120/240 Volts or 110/220 Volts?

Short answer: in the US, 220 volts and 240 volts mean the same thing, and so do 110 and 120. The older numbers just stuck around out of habit.

Utilities in North America supply homes and businesses with 120/240 volts, and the actual voltage can vary by about 5% either way. So a “120-volt” outlet might read anywhere from 114 to 126 volts.

You’ll still see equipment labeled 110/120 (or 220/240), and some manufacturers split the difference with 115 or 230. For all practical purposes, these labels mean the same thing.

Buying for a lab outside North America? Much of the world, including Europe, the UK and Australia, runs on 220-240 volts at 50 Hz instead of 120 volts at 60 Hz. Check that both the voltage and the Hz on the hood’s nameplate match your power. Some hoods are rated for both 50 and 60 Hz, and others come in separate versions.

Not Sure What Your Fume Hood Needs?

National Laboratory Sales will confirm the voltage on any fume hood before you buy, so you know exactly what power to plan for. Browse our fume hoods for sale, read what to ask when purchasing a fume hood, or contact National Laboratory Sales with any questions. If you need an outlet replaced or your lab rewired, call a licensed local electrician.


Fume Hood Voltage FAQ

Can a fume hood plug into a regular outlet?

Most benchtop fume hoods can. They run on a standard 115-120 volt, single-phase outlet. Some hoods and blowers are built for 208-240 volts, so check the nameplate first.

How many amps does a fume hood need?

It depends on the model. Labconco’s Protector hoods, for example, list 10 amps at 115V or 5 amps at 230V. Outlets built into a hood are often rated at 20 amps, and a ducted hood’s blower needs its own power on top of that. Give your electrician the spec sheet so they can size the circuits.

Is a 220V outlet the same as a 240V outlet?

Yes. In the US, 220V and 240V describe the same outlet. 240 is the official number, but people still say 220. The same goes for 110V and 120V.