What is face velocity?

Face velocity is a measurement of the speed at which air enters a fume hood’s face opening. It is generally recommended that a fume hood’s face velocity is between 0.3 m/s (60 fpm) and 0.5 m/s (100 fpm), however, it is best to check with local safety regulations before using a fume hood because variations exist by state and by use.